RESPIRONICS, INC. RESPIRONICS; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
|
Back to Search Results |
|
Model Number V60 V60PLUS VENTILATOR |
Device Problem
Computer Software Problem (1112)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 09/28/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
E1: reporting address state: (b)(6).Reporting address postal: (b)(6).Reporting institution phone #: (b)(6).Reporter phone #: (b)(6).
|
|
Event Description
|
Philips received a complaint by the customer on the v60, indicating that the error for "watchdog test failed" had occurred.The device was in use on a patient at the time the reported issue was discovered; however, there was no harm to the patient or user.A field service engineer (fse) was initially called to the customer's site for a separate issue.At the time of the troubleshoot, the fse discovered that the error "watchdog test failed" had occurred in the event log.The fse then replaced the central processing unit (cpu) printed circuit board assembly (pcba).The fse replaced the cpu pcba to resolve the issue.The device passed required performance verification tests per philips standards and was returned to service.The investigation concludes that no further action is required at this time.If additional information is received the complaint file will be reopened.
|
|