Catalog Number ER420
Device Problems Mechanical Problem (1384); Failure to Form Staple (2579)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/01/2023
Event Type  malfunction  
Event Description
It was reported that during a gastric bypass procedure that the clips were consistently scissoring on the small bowel and gastric staple lines.The clips looked uneven in the jaws prior to ejection.Unknown as to how the procedure was completed.There were no adverse consequences reported.
 
Manufacturer Narrative
(b)(4).Date sent: 12/21/2023.B3: unknown, assumed first day of month that complaint was reported.D4: batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 1/22/2024.D4: batch #x7029v.Investigation summary the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the er420 device was returned with a clip in the jaws and no apparent damage.The clip was removed in order to inspect the jaws and they were found with no damage.The device was tested for functionality.During the analysis, the device was noted with fluids.The device was functional tested, and it fed, retained and formed the remaining eleven(11) clip as intended.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.The event described could not be confirmed as the device performed without any difficulties noted.A manufacturing record evaluation was performed for the finished device batch x7029v number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18372726
MDR Text Key331241219
Report Number3005075853-2023-09737
Device Sequence Number1
Product Code FZP
UDI-Device Identifier10705036012597
UDI-Public10705036012597
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K864102
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberER420
Device Lot NumberX7030K
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 11/22/2023
Initial Date FDA Received12/21/2023
Supplement Dates Manufacturer Received01/04/2024
Supplement Dates FDA Received01/22/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured07/05/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-