Model Number R SERIES |
Device Problem
Electrical Shorting (2926)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Complainant alleged that during biomed testing, the device inappropriately displayed a "defib pad short" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Manufacturer Narrative
|
The device was returned to zoll medical australia; the customer's report was duplicated and attributed to a missing resistor on the pace/defib engine board the pace/defib engine board was replaced to resolve the report.The device was recertified and returned to the customer.Analysis for reports of this type has not identified an increase in trend.
|
|