MEDTRONIC PUERTO RICO OPERATIONS CO. PERCEPT; STIMULATOR, ELECTRICAL, IMPLANTED, FOR PARKINS
|
Back to Search Results |
|
Model Number B35200 |
Device Problems
Break (1069); Failure to Deliver Energy (1211); High impedance (1291); Unintended Collision (1429)
|
Patient Problem
Insufficient Information (4580)
|
Event Date 09/11/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Continuation of d10: product id: b3301542m, lot#serial#: (b)(6), implanted: (b)(6) 2023, explanted: (b)(6) 2023, product type: lead.Section d information references the main component of the system.Other relevant device(s) are: product id: b3301542m, serial/lot #: (b)(6), ubd: 05-oct-2024, udi#: (b)(4).Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Event Description
|
It was reported that patient was not receiving therapy as the impedances were high.Patient mentioned falling around 3 months ago but stated he did not hit his head or neck area, or anywhere near an implant.Patient had battery pocket and lead head incision reopened to test impedances. extension ports were swapped to rule out generator as a cause.The head incision was opened to test the extension lead connection.A lead test cable was used and it showed high impedances on right side.It was also noted that there was a visible break in the lead wires that could not be seen in the x-ray.The surgeon stated it was only visible when the lead was slightly pulled.A revision stage one was performed for the right side and a new lead was placed.
|
|
Manufacturer Narrative
|
H3.Analysis of the lead (s/n (b)(6) found the conductors were broken within 10 centimeters from the proximal end of the connector area.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|