MAUDE Adverse Event Report: RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

RESPIRONICS, INC. BIPAP A40; VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE Back to Search Results
Model Number R1111177
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 02/25/2020
Event Type  malfunction  
Event Description
The manufacturer received information alleging a ventilator inoperable event occurred during use.There was no harm or injury reported.No medical interventions were specified.During the evaluation of the device at the manufacturer's service center, the alleged event could not be confirmed.The device error log has not been recorded properly.The technician ultimately felt it was a software issue due to the current version being v3.6.The technician upgraded the software to v3.6.1 which addresses this failure mode.The technician checked the blower motor and alarm, cleaned the flow line and tested the device on the test station.No abnormalities were found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BIPAP A40
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT, FACILITY USE
Manufacturer (Section D)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer (Section G)
RESPIRONICS, INC.
1001 murry ridge lane
murrysville PA 15668
Manufacturer Contact
kimberly shelly
6501 living place
pittsburgh, PA 15208
4125423300
MDR Report Key18380204
MDR Text Key331221560
Report Number2518422-2023-37226
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K121623
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberR1111177
Device Catalogue NumberR1111177
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/11/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 02/25/2020
Initial Date FDA Received12/21/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/29/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-