Catalog Number 01.25.021
Device Problem Device Dislodged or Dislocated (2923)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 11/17/2023
Event Type  Injury  
Manufacturer Narrative
Batch review performed on 13 december 2023: lot 2308541: (b)(4) items manufactured and released on 24-may-2023.Expiration date: 2028-05-09.No anomalies found related to the problem.To date, 31 items of the same lot have been sold without any similar reported event in the period of review.Other device involved: liner: mpact 01.32.3239hct flat pe hc liner 32/c (k103721) lot 2309697: (b)(4) items manufactured and released on 13-june-2023.Expiration date: 2028-09-20.No anomalies found related to the problem.To date, 39 items of the same lot have been sold without any similar reported event in the period of review.
 
Event Description
Hip instability due to a dislocation of the head from the liner.About 1 month post primary the surgeon revised the medacta head with another medacta head and the medacta cup and liner were revised to a competitor's product.The surgery was completed successfully.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BALL HEADS: COCR 01.25.021 COCR BALL HEAD 12/14 Ø 32 SIZE S -3.5
Type of Device
HIP COCR BALL HEAD
Manufacturer (Section D)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ  6874
Manufacturer (Section G)
MEDACTA INTERNATIONAL SA
strada regina
castel san pietro, switzerland 6874
SZ   6874
Manufacturer Contact
marco giannessi
strada regina
castel san pietro, switzerland 6874
SZ   6874
MDR Report Key18380308
MDR Text Key331177590
Report Number3005180920-2023-01046
Device Sequence Number1
Product Code JDI
UDI-Device Identifier07630030805158
UDI-Public07630030805158
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K072857
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 12/21/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number01.25.021
Device Lot Number2308541
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/01/2023
Initial Date FDA Received12/21/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured03/24/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-