The device was not returned to zoll medical corporation for evaluation.Instead, the clinical file of the customer's report was provided.Review of the clinical file did show the customer's report.The waveform measurements in segments 2 and 3 fell within the range of measurements defined for vf.The percentages of detected peaks with similar amplitudes were 64% and 65% for segments 2 and 3 where the defined upper limit is less than 66%.The number of peaks detected in segments 2 and 3 was 11 for each segment and the algorithm matching definition is looking for a value of greater than 10.In each instance, these measurements are borderline and resulted in a false positive result of shock advised.The aed operated as designed and within the limitations of the technology.Analysis of reports of this type has not identified an increase in trend.
|