It was reported that the cable for disposables (hls cable) was defective which resulted in wrong pressure value readings.The failure occurred during priming.A getinge field service technician (fst) was sent for investigation and repair.The failure could be traced back to the connection cable for disposables, which was replaced.The fst performed safety, calibration, and functionality checks to factory specifications.All function tests are passed.The exact root cause remains unknown.However, similar failures with a variety of causes were investigated in the past: another disposable connection cable with a similar failure was investigated by getinge life cycle engineering on 2021-03-31.Deposit and oxides were found on the cable socket.By wetting the socket plane with salt-containing liquids (priming), the measurement signals were influenced by the electrical conductivity of the contamination.A service bulletin (issue 95 / 21-05-04) was published may 2021 to make the users aware that the contacts of the plug connections must not come into contact with cleaning agents, disinfectants or priming liquid.Additionally, another disposable connection cable with a similar failure was investigated by getinge life cycle engineering (lce) on 2022-11-02.The nature of the error could be traced back to a missing electrical connection within the cable.The root cause for the missing connection is a broken wire within the cable, which is originated from external force.According to the instruction for use of the involved disposables (hls set advanced 5.0 / 7.0, hit set advanced 5.0 / 7.0, v2.4, chapter 6.1 preparation and installation and quadrox-ir small adult / adult, chapter 7.2 priming the system) the pressure sensors have to be calibrated and checked before priming.Furthermore, the cardiohelp has a flow/bubble sensor and a venous probe to measure and control the blood flow and parameters.If the measured values are above high limit or below low limit of the set limits the system generates a visual and acoustical alarm.Further, it is stated in the instruction for use (ifu) of the cardiohelp, chapter 5.3 "connection the sensors", to ensure that the connected sensors are not defective and to not use, if there is a visible damage.The review of the non-conformities has been performed on 2023-12-18 for the period of 2016-08-19 to 2023-11-13.It does not show any non-conformity in regard to the reported product and failure.There is no indication that manufacturing issues occurred during this time, thus production related influences are unlikely.The device was manufactured on 2016-08-19.Based on the results the reported failure "pressure readings wrong due to hls cable" could be confirmed.The customer will be informed about the results by the getinge sales and service unit.The occurrence rate was calculated for the reported issue and it was determined that this is not a systemic issue.Therefore, no remedial action is required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿ s trending program and additional investigations or corrections will be implemented in case of adverse trending.
|