• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET SAS HLED; LIGHT, SURGICAL, CEILING MOUNTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET SAS HLED; LIGHT, SURGICAL, CEILING MOUNTED Back to Search Results
Model Number ARD568370953
Device Problem Leak/Splash (1354)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/26/2023
Event Type  malfunction  
Manufacturer Narrative
Event site name: event site city: additional information will be provided following the conclusion of the investigation.H3 other text : device not returned to manufacturer.
 
Event Description
On 26th november, 2023 getinge became aware of an issue with one of surgical lights - hled.It was stated there are oil droplets under the stainless steel dust cover of the spring arm, with the risk of dripping off.We decided to report the issue in abundance of caution as any liquid falling off into sterile field or during procedure may cause contamination.
 
Manufacturer Narrative
The correction of h3a device evaluated by mfg, h3b device not eval provide code and h3c if other provide code-explain deems required.This is based on the internal evaluation.Previous h3a device evaluated by mfg: no.Corrected h3a device evaluated by mfg: yes.Previous h3b device not eval provide code: other.Corrected h3b device not eval provide code: none.Previous h3c device not eval provide code: device not returned to manufacturer.Corrected h3c device not eval provide code: none.Getinge became aware of an issue with one of surgical lights - hled.It was stated there are oil droplets under the stainless steel dust cover of the spring arm, with the risk of dripping off.We decided to report the issue in abundance of caution as any liquid falling off into sterile field or during procedure may cause contamination.It was established that when the event occurred, the surgical light did not meet its specification due to oil leakage from hled surgical light, which contributed to the event.Provided information does not indicate if upon the event occurrence, the device was or was not being used for patient treatment.During the investigation, it was found that in the past the reported scenario has never led to serious injury nor death.Comparing the number of involved devices to the install base, we conclude that the failure ratio for the allegation of liquid leakage is low.As stated by the subject matter expert at the manufacturer¿s site, during the assembly of spring arms the supplier applies a creamed grease.Such creamed grease is turning into liquid only above 135°c.So, the black dripping liquid observed, therefore, cannot come from the spring arm itself but finds its origin elsewhere.The first likely cause is a combination of air conditioning and an excess of oil at the bushing location between the spring arm and the main arm.And according to the latest technical investigations carried out in august 2020 at maquet sas the second probable root cause would be an excess of cleaning product in the lower part of the spring arm, applied in spray or with a sponge, but not in compliance with the recommendations given in our user manual.Manufacturer¿s recommends following cleaning protocol described in user manual (hled um 01601en08_extract, pages 34-35).Getinge shall continue to monitor for any further events of this nature and do not propose any further action at this time.
 
Event Description
Manufacturer's reference number: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HLED
Type of Device
LIGHT, SURGICAL, CEILING MOUNTED
Manufacturer (Section D)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer (Section G)
MAQUET SAS
parc de limere
avenue de la pomme de pin
ardon
Manufacturer Contact
pascal jay
parc de limere
avenue de la pomme de pin
ardon 
MDR Report Key18382695
MDR Text Key331192094
Report Number9710055-2023-00982
Device Sequence Number1
Product Code FSY
Combination Product (y/n)N
Reporter Country CodeTH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberARD568370953
Device Catalogue NumberARD568370953
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/26/2023
Initial Date FDA Received12/22/2023
Supplement Dates Manufacturer Received04/26/2024
Supplement Dates FDA Received04/29/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2014
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-