ST. JUDE MEDICAL TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number A-TFSE-DF
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cardiac Perforation (2513)
Event Date 12/14/2023
Event Type  Injury  
Event Description
During a ventricular tachycardia ablation procedure, a pericardial effusion occurred which required a pericardiocentesis.Ablation had been performed in the left ventricle and towards the end of the procedure, the patient became hypotensive.A transthoracic echocardiogram revealed a pericardial effusion for which a pericardiocentesis was performed to stabilize the patient.There were no performance issues with any abbott devices.
 
Manufacturer Narrative
The results of the investigation are inconclusive since the device was not returned for analysis.Further information regarding the procedure was obtained stating the power settings used during the procedure was between 40-50 watts, and point-by-point ablations were performed lasting up to 60 seconds.The tactiflex ablation catheter sensor enabled instructions for use (ifu) recommends ablation duration between 10-20 seconds for power settings between 40-50 watts.Due to unknown procedural conditions, we are unable to conclusively determine the cause of the reported pericardial effusion.Per the ifu, cardiac tamponade and pericardial effusion are known risks during the use of this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TACTIFLEX¿ ABLATION CATHETER, SENSOR ENABLED¿
Type of Device
CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
Manufacturer (Section G)
ST. JUDE MEDICAL
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS   1897-4050
Manufacturer Contact
janna parks
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key18382741
MDR Text Key331193815
Report Number3008452825-2023-00593
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
P220013
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberA-TFSE-DF
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received12/22/2023
Supplement Dates Manufacturer Received12/28/2023
Supplement Dates FDA Received01/02/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-