• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON DICKINSON CARIBE LTD. BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON DICKINSON CARIBE LTD. BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC); SYSTEM, BLOOD CULTURING Back to Search Results
Catalog Number 442021
Device Problem False Negative Result (1225)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/07/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.D4.Medical device lot#: unknown.D4.Medical device expiration date: unknown.H4.Device manufacture date: unknown.
 
Event Description
It was reported when using bd bactec¿ lytic/10 anaerobic/f culture vials (plastic) the customer had false negative results.There was no health impact or consequences reported.
 
Manufacturer Narrative
H.6 investigation summary catalog 442023; batch no.Unknown.Customer reported a false negative result.Two photos of bactec bottles were received.Bd was not able to perform an investigation to the retention samples since batch number is unknown.A complaint history review cannot be conducted relating to the incident lot number and the ¿as reported¿ defect code since batch number is unknown.The batch history record could not be reviewed as the lot number is unknown nonetheless batch history records are always reviewed prior to product release.Users are cautioned in the package insert under limitation of the procedure: a gram-stained smear from a culture medium may contain small numbers of nonviable organisms derived from media constituents, staining reagents, immersion oil, glass slides, and specimens used for inoculation.In addition, the patient specimen may contain organisms that will not grow in the culture medium or in media used for subculture.Such specimens should be subculture to special media as appropriate.Complaint is unconfirmed.No correctives actions were required.A cross functional team continually monitors all product complaints for trends and determines if any additional actions are necessary beyond the current investigation process.
 
Event Description
It was reported when using bd bactec¿ lytic/10 anaerobic/f culture vials (plastic) the customer had false negative results.There was no health impact or consequences reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD BACTEC¿ LYTIC/10 ANAEROBIC/F CULTURE VIALS (PLASTIC)
Type of Device
SYSTEM, BLOOD CULTURING
Manufacturer (Section D)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer (Section G)
BECTON DICKINSON CARIBE LTD.
vicks drive
lot no. 6
cayey PR
Manufacturer Contact
jo doyka
7 loveton circle
sparks, MD 21152
4103164000
MDR Report Key18385646
MDR Text Key331489373
Report Number2647876-2023-00668
Device Sequence Number1
Product Code MDB
UDI-Device Identifier00382904420215
UDI-Public(01)00382904420215
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K123903
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number442021
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/28/2023
Initial Date FDA Received12/22/2023
Supplement Dates Manufacturer Received02/05/2024
Supplement Dates FDA Received02/20/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-