The device was returned for evaluation.Inspection and testing found that during start-up there was a smoking smell and sparks due to a faulty ac/dc power supply.Additionally, the exhaust fan in the ventilation system was not functioning properly, there was dust build up and foreign material inside the fan and power supply.Based on the facts obtained in the investigation, it is likely that dust had accumulated in the power supply unit and fan, which temporarily ignited the dust when the power was turned on, causing sparks and smoke at the same time.From this, it is presumed that the power supply unit and fan failed due to the influence of dust.However, a definitive root cause could not be determined.A review of the device history record found no deviations that could have caused or contributed to the reported issue.It has been over 8 years since the subject device was manufactured.Olympus will continue to monitor field performance for this device.
|