MAUDE Adverse Event Report: WRIGHT MEDICAL TECHNOLOGY INC UNKNOWN TIBIAL COMPONENT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WRIGHT MEDICAL TECHNOLOGY INC UNKNOWN TIBIAL COMPONENT; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER Back to Search Results
Catalog Number UNK_WWA
Device Problems Break (1069); Fracture (1260); Osseointegration Problem (3003)
Patient Problems Cyst(s) (1800); Inadequate Osseointegration (2646); Insufficient Information (4580)
Event Date 11/29/2023
Event Type  Injury  
Manufacturer Narrative
The device is not available for evaluation as it remains implanted in the patient.A review of the device history is not possible because the lot number was not communicated.If additional information becomes available, it will be provided on a supplemental report.H3 other text : device remains implanted in patient.
 
Event Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to lateral tibial peg broken on tibial implant.
 
Manufacturer Narrative
Correction - h6 (clinical signs code, device code, results code and conclusion code).The complaint could be confirmed, since the information for evaluation matches the alleged failure, as a revision could be confirmed based on implant loosening, but a implant breakage could not be confirmed based on the received information.Upon further investigation of the ct scans by healthcare professionals the following was observed: there is clear radiolucence and smaller anterior cysts.These findings suggest loosening.There is no clear sign of migration visible.The implant is in an intact condition.A review of the device history was not possible because the lot number was not communicated.No corrective actions are required at this time.A review of the labeling did not indicate any abnormalities.Indications of material, manufacturing, or design related problems were unable to be identified as the lot number was not communicated.Based on investigation, the root cause was attributed to a patient factors issue.If the device is returned or if any additional information is provided, the investigation will be reassessed.
 
Event Description
The digital stryker prophecy team received a ct scan indicating that the patient may require a revision surgery due to lateral tibial peg broken on tibial implant.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN TIBIAL COMPONENT
Type of Device
PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
Manufacturer (Section D)
WRIGHT MEDICAL TECHNOLOGY INC
1023 cherry rd
memphis TN 38117
Manufacturer (Section G)
WRIGHT MEDICAL TECHNOLOGY, INC.
11576 memphis arlington rd
arlington TN 38002
Manufacturer Contact
anna jusinski
325 corporate drive
mahwah, NJ 07430
2018315000
MDR Report Key18387313
MDR Text Key331288223
Report Number3010667733-2023-00807
Device Sequence Number1
Product Code HSN
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/28/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Catalogue NumberUNK_WWA
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2023
Initial Date FDA Received12/22/2023
Supplement Dates Manufacturer Received02/02/2024
Supplement Dates FDA Received02/28/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient SexMale
-
-