LIVANOVA DEUTSCHLAND GMBH SCP DRIVE UNIT W/FAST CLAMP; CONTROL, PUMP SPEED, CARDIOPULMONARY BYPASS
|
Back to Search Results |
|
Model Number 60-01-00 |
Device Problem
Pumping Problem (3016)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/29/2023 |
Event Type
malfunction
|
Event Description
|
Livanova deutschland received a report that, during a procedure, the scp pump keeps resetting the rpm to 1000 and displays error eprom defective (e04).There is no report of any patient injury.
|
|
Manufacturer Narrative
|
A.1.-a.5.Patient information was not provided.H10: livanova deutschland manufactures the cp5.The incident occurred in usa.Unit was inspected by field service engeneed and was i found that the control panel was not fully plugged into the console base.When connection was corrected, everything started working properly.Unlikely that the issue was caused by a not fully plugges control panel if any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
|
|
Event Description
|
See initial report.
|
|
Manufacturer Narrative
|
The error code e04 indicates that eprom is defective and the involved component is the cpu board.The complaints database review revealed that no similar issues have been submitted for this unit since its installation in 2017.Based on the above facts, the most likely root cause is an electrical failure, as bad/loose connections, a false contact or some oxidized electronics on the board that was fixed during checks performed by the fse during inspection.Failure of electronic boards can be related to multiple and not deterministic factors such as exposure to heat, dust and moisture, accidental impacts (drops and falls), and power overloads/surges but also to variability of micro subcomponents.However, there is no concerning trend for this kind of failure.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
|
|
Search Alerts/Recalls
|
|
|