• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. AORTIC AP; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. AORTIC AP; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 505DA20
Device Problem Detachment of Device or Device Component (2907)
Patient Problem Insufficient Information (4580)
Event Date 05/02/2023
Event Type  malfunction  
Manufacturer Narrative
Product analysis: upon receipt at medtronic's quality laboratory, visual examination revealed that there was evidence of blood contact on the returned device.One leaflet was returned with the valve.The attached leaflet appeared intact; surface anomalies (scratch-like marks) were found on the outflow surface.Similar surface anomalies were found on the detached leaflet.Coaptation of the attached leaflet was tested.The attached leaflet was fully mobile.Orifice appeared intact; surface anomalies (scratch-like marks) were found on the orifice where detached leaflet was located.The hinge mechanisms were intact with no evidence of damages.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Event Description
Medtronic received information that during the implant of this 20mm aortic mechanical valve, it was reported that one leaflet came off after the valve was being "tightened to ligate the valve ring".It was stated that the physician believes this was not due to an external pressure on the valve but by pressure on the leaflet when checking the tissue under the valve by suctioning.It was reported that the leaflet did not split or smash and there were no remains in the body of the patient.The valve was replaced with a mechanical valve of the same size and model.No additional adverse patient effects were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AORTIC AP
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer (Section G)
MEDTRONIC, INC.
3800 annapolis lane
minneapolis MN 55447
Manufacturer Contact
alison sweeney
parkmore business park west
galway 
EI  
091708096
MDR Report Key18388945
MDR Text Key331300408
Report Number3008592544-2023-00114
Device Sequence Number1
Product Code LWQ
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P990046
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/22/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number505DA20
Device Catalogue Number505DA20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/23/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 05/07/2023
Initial Date FDA Received12/22/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/12/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-