The product was returned to pentax medical for repair.Our technician checked the returned unit and confirmed that the ccd module with drive pcb laser filter lifting.Based on the result, we concluded that it was caused due to the excessive force applied on the ccd module with drive pcb.In addition, our technician confirmed that the lcb (light carrying bundle) broken, the ccd module with drive pcb cloudy (not clear view), and the u/d pulley wire worn out; however, these defects are not the main cause, and/or irrelevant to the alleged complaint.Based on the technical report ""hr-rpt-0586(image failure)"" and/or the risk analysis results, it was evaluated to submit mdr.
|