• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - ALCON PRECISION DEVICE ALCON CLOSURE SYSTEM (ACS), NYLON; SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE Back to Search Results
Catalog Number 8065198001
Device Problems Break (1069); Dull, Blunt (2407)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/01/2023
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that an ophthalmic operating suture was broken and the needle does not cut during the surgery.Surgery was completed on the same day and patient impact was not reported.
 
Manufacturer Narrative
The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
One unopened 10/0 bmn suture in a blister in a pouch was received for the report of breakage of nylon, needle does not cut.Sample was visually inspected and found to be conforming.A dimensional inspection was also done for suture diameter and found to be conforming.Functional testing for suture strength was performed and found to be conforming.A review of the device history record traceable to the reported lot number, indicates that the reported product was processed and released according to the reported product¿s acceptance criteria.The photo attached to the parent file was reviewed by the manufacturing site.Photo show package with reported lot number.Reported issue cannot be confirmed from photo.The returned unopened sample was found to be conforming for all visual, dimensional, and functional testing associated with the reported issue, therefore the breakage of nylon, needle does not cut as described in the complaint was not confirmed.However since the actual complaint samples were not returned, a root cause cannot be determined for the complaint as described by the customer.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.Suture material samples are tested at incoming for tensile strength and diameter and visually inspected for any defects such as discoloration and frays.Sutures are also 100% inspected by trained operators for discoloration and frays and for proper attachment during the manufacturing process.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No adverse trends have been observed associated with the reported product and event.No additional action is required at this time.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALCON CLOSURE SYSTEM (ACS), NYLON
Type of Device
SUTURE, NONABSORBABLE, SYNTHETIC, POLYAMIDE
Manufacturer (Section D)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer (Section G)
ALCON RESEARCH, LLC - ALCON PRECISION DEVICE
714 columbia avenue
sinking spring PA 19608
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18391336
MDR Text Key331292911
Report Number2523835-2023-00746
Device Sequence Number1
Product Code GAR
UDI-Device Identifier00380651980013
UDI-Public00380651980013
Combination Product (y/n)N
Reporter Country CodeSP
PMA/PMN Number
K930517
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup
Report Date 05/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number8065198001
Device Lot Number155LL7
Was Device Available for Evaluation? Device Returned to Manufacturer
Initial Date Manufacturer Received 12/14/2023
Initial Date FDA Received12/26/2023
Supplement Dates Manufacturer Received01/24/2024
04/25/2024
Supplement Dates FDA Received02/20/2024
05/22/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/03/2022
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-