• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS MX40 1.4 GHZ SMART HOPPING Back to Search Results
Model Number 865350
Device Problem No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/11/2023
Event Type  malfunction  
Manufacturer Narrative
Diagnostic/functional testing was performed at the philips authorized repair facility.Results of functional testing indicate that there was no speaker sound during the start up test.The device failed a manual power on test, and speaker was defective.The cause of the reported problem was confirmed to be a defective speaker.The speaker was replaced.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.
 
Event Description
During evaluation at philips bench repair, it was identified that the device had no audio.The device was not in clinical use at the time the issue was discovered.No adverse event or harm was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MX40 1.4 GHZ SMART HOPPING
Type of Device
MX40 1.4 GHZ SMART HOPPING
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18393206
MDR Text Key331548368
Report Number1218950-2023-00970
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113125
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number865350
Device Catalogue Number865350
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/11/2023
Initial Date Manufacturer Received 12/11/2023
Initial Date FDA Received12/26/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/16/2021
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
-
-