• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITHS MEDICAL ASD, INC. TUBES; SET, ADMINISTRATION, INTRAVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITHS MEDICAL ASD, INC. TUBES; SET, ADMINISTRATION, INTRAVASCULAR Back to Search Results
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/14/2023
Event Type  malfunction  
Event Description
Patient reported with last 2 monthly supply shipments all of the cassettes lot 4396110 have 2 metal staples holding the cassette tubing down on top of cassette placed 1/2 in apart instead of the single plastic bar between the 2 green plastic pieces that are under the blue clip when removed.Pt reported hi pressure alarms with about 10 out of 30 cassettes that resolved when pt removed the 2 staples.Also, the 60 inch ext tubing is not packaged as securely coiled to prevent kinks and there are multiple kinks in the tubing lot 4387695 causing increased hi pressure alarms that pt is able to resolve by manually smoothing the tubing kinks.This has occurred in about 8 out 30 tubing¿s.Smith cadd 100 ml flow stop cassettes and smiths 60 in ext tubing.Reference reports: #mw5149511, #mw5149512, #mw5149513, #mw5149514, #mw5149515, #mw5149516, #mw5149517, #mw5149518, #mw5149519, #mw5149520, #mw5149521, #mw5149522, #mw5149523, #mw5149524, #mw5149525, #mw5149527, #mw5149528.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TUBES
Type of Device
SET, ADMINISTRATION, INTRAVASCULAR
Manufacturer (Section D)
SMITHS MEDICAL ASD, INC.
MDR Report Key18393291
MDR Text Key331483541
Report NumberMW5149526
Device Sequence Number1
Product Code FPA
Combination Product (y/n)Y
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Pharmacist
Type of Report Initial
Report Date 12/14/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received12/22/2023
Patient Sequence Number1
Patient Age60 YR
Patient SexMale
Patient Weight64 KG
Patient EthnicityNon Hispanic
-
-