Model Number URF-V2 |
Device Problem
Poor Quality Image (1408)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was returned to olympus for evaluation.In addition to the reportable malfunction documented in b5, the evaluation findings are as follows: the control unit was sticky due to water leakage; the universal cord and the up/down angulation lever plate was sticky; the light guide bundle was slipping down; the universal cord had a scratch; the adhesive on the bending section cover had a chip and due to a pinhole on the bending section cover, water tightness was lost.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
|
|
Event Description
|
The customer reported to olympus, the uretero-reno videoscope had horizontal lines like a sandstorm appear on the screen.No patient harm reported.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: due to corrosion on the angle mechanism, the bending section could not be controlled at all.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
|
Manufacturer Narrative
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the observed angulation issue could not be determined, however, the issue was likely the result of observed corrosion of the angulation mechanism.The event may be prevented by following the instructions for use section below: chapter 3 preparation and inspection 3.3 inspection of the endoscope.Olympus will continue to monitor field performance for this device.
|
|
Search Alerts/Recalls
|
|
|