AIZU OLYMPUS CO., LTD. URETERO-RENO VIDEOSCOPE Back to Search Results
Model Number URF-V2
Device Problem Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation.In addition to the reportable malfunction documented in b5, the evaluation findings are as follows: the control unit was sticky due to water leakage; the universal cord and the up/down angulation lever plate was sticky; the light guide bundle was slipping down; the universal cord had a scratch; the adhesive on the bending section cover had a chip and due to a pinhole on the bending section cover, water tightness was lost.The investigation is ongoing and follow up with the user facility is currently being performed.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the uretero-reno videoscope had horizontal lines like a sandstorm appear on the screen.No patient harm reported.The device was returned for evaluation.During the device evaluation, the following reportable malfunction was found: due to corrosion on the angle mechanism, the bending section could not be controlled at all.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause of the observed angulation issue could not be determined, however, the issue was likely the result of observed corrosion of the angulation mechanism.The event may be prevented by following the instructions for use section below: chapter 3 preparation and inspection 3.3 inspection of the endoscope.Olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO VIDEOSCOPE
Type of Device
URETERO-RENO VIDEOSCOPE
Manufacturer (Section D)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima 965-8 520
JA  965-8520
Manufacturer (Section G)
AIZU OLYMPUS CO., LTD.
3-1-1 niiderakita
aizuwakamatsu-shi, fukushima
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18394391
MDR Text Key331373796
Report Number9610595-2023-20316
Device Sequence Number1
Product Code FGB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K172246
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/06/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-V2
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/29/2023
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/18/2023
Initial Date FDA Received12/26/2023
Supplement Dates Manufacturer Received01/30/2024
Supplement Dates FDA Received02/06/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/17/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-