• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MALE LL ADAPTOR; INTRAVASCULAR ADMINISTRATION SET

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SISTEMAS MEDICOS ALARIS S.A. DE C.V. BD MALE LL ADAPTOR; INTRAVASCULAR ADMINISTRATION SET Back to Search Results
Catalog Number 1001-110-004
Device Problem Manufacturing, Packaging or Shipping Problem (2975)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 11/29/2023
Event Type  malfunction  
Manufacturer Narrative
H.3.If a device evaluation and/or device history review is completed, a supplemental report will be filed.
 
Event Description
It was reported that bd male ll adaptor had mixed products the following information was received by the initial reporter with the verbatim: 1001-110-004 were mixed with other types of products.
 
Manufacturer Narrative
The customer reported that 1001-110-004 were mixed with other types of products.There were pictures provided of the mixed product, and upon visual inspection, the complaint was verified.The manufacturing site, north america molding center (namc), found the root cause to be human error in placing the materials into the wrong bin.Namc implemented corrective and preventative actions to fix the issue.This involved adding arrows on the floor for operators, updating the line clearance form, retaining and updating the itr for associates, and adding signs above the bins that match the press number.Namc performed a device history record's review (dhr) including a review of all data collected during in process and quality inspections.The batches involved in this complaint meet all acceptable quality levels (aql's), were manufactured, and released according to applicable procedures and specifications.Quality notification query for the mixed product condition was performed, and there was no history of the defect from batch 2291702.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.Your assistance in this matter has been helpful in trend identification and supporting our commitment to continuous quality improvement.
 
Event Description
No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD MALE LL ADAPTOR
Type of Device
INTRAVASCULAR ADMINISTRATION SET
Manufacturer (Section D)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer (Section G)
SISTEMAS MEDICOS ALARIS S.A. DE C.V.
blvd. insurgentes no. 20351
parque industrial el florido
tijuana
Manufacturer Contact
helen cox (mdr)
75 north fairway drive
vernon hills, IL 60061
8473935694
MDR Report Key18395231
MDR Text Key331676495
Report Number9616066-2023-02438
Device Sequence Number1
Product Code FPA
UDI-Device Identifier70885403483778
UDI-Public(01)70885403483778
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number1001-110-004
Device Lot NumberUNKNOWN
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/29/2023
Initial Date FDA Received12/26/2023
Supplement Dates Manufacturer Received01/04/2024
Supplement Dates FDA Received01/04/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-