EXACTECH, INC. NV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS; PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED Back to Search Results
Model Number NV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS
Device Problem Naturally Worn (2988)
Patient Problem Failure of Implant (1924)
Event Date 12/01/2023
Event Type  Injury  
Event Description
As reported by the germany competent authorities and as part of the manufacturer's recall campaign, the patient went to have the device implanted in 2019 checked left hip prosthesis.The x-ray control showed a clear decentering of the prosthetic head and unusually large osteolysis in the acetabulum as a sign of inlay wear.This could happen to one inlay change on (b)(6) 2023 to a vitd-hardened, specially approved inlay novation xle neutral liner, group 1, 32mm (cat# 140-32-51 / serial# (b)(6)).As part of the replacement operation, the inlay exchange, determination of solid cup integrity as well as curettage and sealing of the cysts socket bearing using allogeneic cancellous bone and changing the prosthetic head.The explants are currently in the laboratory doctor's office for microbiological examination using sonication.Surgical reports/x-rays are also available upon request.Hospital data of the implantation and replacement, as well as photos of the explants, will be sent upon receipt.
 
Manufacturer Narrative
Section d10: concomitant products: crown cup, cluster-hole gr.50 (cat# 180-01-50 / serial# (b)(6)).Additional information, including the product investigation, will be submitted within 30 days of receipt.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer (Section G)
EXACTECH, INC.
2320 nw 66 court
gainesville FL 32653
Manufacturer Contact
miguel soza
MDR Report Key18395374
MDR Text Key331359366
Report Number1038671-2023-03058
Device Sequence Number1
Product Code JDI
UDI-Device Identifier10885862207074
UDI-Public10885862207074
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K121392
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Remedial Action Recall
Type of Report Initial
Report Date 12/26/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/26/2023
Device Model NumberNV GXL LINER NEUTRAL, 32MM ID GROUP 1 CUPS
Device Catalogue Number130-32-51
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2023
Initial Date FDA Received12/26/2023
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/30/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction NumberZ-1732-2022
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient SexFemale
-
-