Model Number BML-V442QR-30
Device Problem Material Split, Cut or Torn (4008)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/12/2023
Event Type  malfunction  
Manufacturer Narrative
The device was returned to olympus for evaluation and the customer's allegation of the tip was torn was confirmed.The device evaluation found the guidewire tip was torn with no parts missing.The investigation is ongoing.A supplemental report will be submitted upon completion of the investigation or if any additional information is provided by the user facility.
 
Event Description
The customer reported to olympus, the single use mechanical lithotriptor tip was torn to pieces before use.The issue was found during pre-use inspection for a therapeutic endoscopic retrograde cholangiopancreatography (ercp) procedure.Another similar device was used to complete the procedure there were no reports of patient harm.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation the specific cause of the torn tip was not established however, there are multiple factors that may have contributed to the reported issue including: 1) an attempt was made to insert the device into the endoscope while using the guide wire.2) when the angle between the distal end and the guide wire was large as shown in the following figure, the device was inserted into the endoscope.3) a force exceeding the resisting force was applied to the guide wire tip.This caused the guide wire tip to tear.The occurrence of the reported problem can be prevented by adhering to the instructions for use (ifu) which states the following: ¿when using the guide wire, insert the instrument with its distal tip in parallel with the guide wire while holding the distal tip as shown in figure 4.20.Be careful not to forcibly insert the instrument with a sharp angle between the distal tip and the guide wire as shown in figure 4.21.This may damage the distal tip.¿ olympus will continue to monitor field performance for this device.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE MECHANICAL LITHOTRIPTOR V
Type of Device
SINGLE USE MECHANICAL LITHOTRIPTOR
Manufacturer (Section D)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori 036-0 357
JA  036-0357
Manufacturer (Section G)
AOMORI OLYMPUS CO., LTD.
2-248-1 okkonoki
kuroishi-shi, aomori
Manufacturer Contact
todd brill
800 west park drive
westborough, MA 01581
5082077661
MDR Report Key18397722
MDR Text Key331716862
Report Number9614641-2023-01995
Device Sequence Number1
Product Code LQC
UDI-Device Identifier04953170218422
UDI-Public04953170218422
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
CLASS2-EXMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberBML-V442QR-30
Device Lot Number3XK
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/21/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received12/27/2023
Supplement Dates Manufacturer Received01/18/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/04/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial