• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION AGILE ESOPHAGEAL; PROSTHESIS, ESOPHAGEAL Back to Search Results
Model Number M00517460
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  Injury  
Manufacturer Narrative
Block h6: imdrf device code a150103 captures the reportable event of stent prematurely deployed.
 
Event Description
It was reported to boston scientific corporation that an agile esophageal fully covered stent was to be implanted in the esophagus to treat an esophageal stricture due to esophageal cancer during an esophagogastroduodenoscopy (egd) with stent placement procedure performed on (b)(6) 2023.The patient's anatomy was tortuous and was not dilated prior to stent placement.During the procedure, the stent prematurely deployed.The stent was removed using rat tooth forceps and another agile esophageal stent was used to complete the procedure.There were no patient complications as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AGILE ESOPHAGEAL
Type of Device
PROSTHESIS, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
ballybrit business park
galway
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key18399113
MDR Text Key331402070
Report Number3005099803-2023-06884
Device Sequence Number1
Product Code ESW
UDI-Device Identifier08714729973089
UDI-Public08714729973089
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K180144
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial
Report Date 12/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberM00517460
Device Catalogue Number1746
Device Lot Number0031327572
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/08/2023
Initial Date FDA Received12/27/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured03/29/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-