Catalog Number MCS20 |
Device Problems
Difficult to Open or Close (2921); Mechanical Jam (2983)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 11/01/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Date sent: 12/27/2023.B3: unknown, assumed first day of month that complaint was reported.D4: batch # unk.Attempts have been made to retrieve the device.To date the device has not been returned.If the device or further details are received at a later date a supplemental medwatch will be sent.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
|
|
Event Description
|
It was reported that during an unknown procedure, the jaws did not open, and the clip could not be fed into the jaw, although the clips were remained.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|
|
Manufacturer Narrative
|
(b)(4).Date sent: 1/31/2024.D4: batch # a9d964.Investigation summary: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcs20 device was returned with no damage to the external components.In an attempt to replicate the reported event, the device was tested for functionality.Upon testing, the device was fired, and the clips did not advance into the jaw.The device was disassembled to evaluate the condition of the internal components and the feedbar was found damaged, causing the firing issue.Six(6) clips were found inside the clip track.Although no conclusion could be reached on what caused the damage to the feed bar, it is possible that the device was clipped over a hard object resulting in higher force to fire, subsequently damaging the mechanism.Please note that the instructions for use contain the following caution: ensure that a clip is in the jaws.Position the jaws so that the clip completely surrounds the vessel to be ligated.Fully squeeze the handles together until they stop.As the handles are squeezed, the clip closes and occludes the vessel or tubular structure.Fully release the handles to return the instrument to its original open position.The next clip is automatically advanced.Inspect the jaws after each clip application to ensure the presence of a new clip before applying the next clip.Repeat steps as necessary to continue ligation of tissue.For optimal performance, periodically clean the instrument during the procedure by submerging the jaw and distal end of the applier shaft in saline.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch number, and no non-conformances were identified.
|
|
Search Alerts/Recalls
|
|
|