PHILIPS MEDICAL SYSTEMS PATIENT INFORMATION CENTER IX Back to Search Results
Model Number 866389
Device Problem Defective Alarm (1014)
Patient Problem Cardiac Arrest (1762)
Event Date 11/24/2023
Event Type  Death  
Manufacturer Narrative
A follow up report will be submitted once the investigation is complete.A philips technical consultant (tc) went to the customer¿s site and checked the mx40 logs, event logs, and pic ix configuration.The tc verified alarms on mx40 properly worked and sent the customer feedback.Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.
 
Event Description
The customer reported the unit did not alarm on (b)(6) 2023 at 0905, a code blue was called after the patient was found unresponsive by the licensed practical nurse (lpn).When reviewing the 20-minute telemetry disclosure, it was found that the patient had been asystole for 20 minutes prior to the code being called.The patient¿s pacemaker was spiking but there was no underlying rhythm.The central monitoring station did not alarm asystole until (b)(6) 2023 09:04.The patient passed away.The mx40 telemetry device in use during this event is reported in mfr number 1218950-2023-00955.
 
Manufacturer Narrative
A philips remote service engineer (rse) remotely interviewed the customer who was onsite.The rse decided to dispatch a philips technical consultant (tc) onsite to further investigate and to collect the logs to be reviewed by an internal philips product support engineer (pse).The tc verified alarms activated whenever simulated thresholds exceeded.He found there were no issues.The internal philips product support engineer (pse) examined the same logs reviewed by the tc and found that 7 north bed label 725b was the one most frequently indicated in the alarm logs during the time period in question, first indicating the asystole at 8:23:08 on (b)(6), roughly 40 minutes prior to the customer assertion.Bed 725b shows multiple asystole alarms over the time period of concern, each looks to have been silenced at the central.The device remains at the customer site.No further investigation or action is warranted at this time.
 
Event Description
This report is based on information provided by our philips complaint handling team.An 85 year old male was admitted on (b)(6) 2023 for low hemoglobin in the setting of dyspnea on exertion (doe) due to diastolic congestive heart failure (chf) and aortic insufficiency.The patient was being followed by the heart failure team and was transitioned from intravenous piggyback therapy (ivpb).The patient received 1 unit of packed red blood cells (prbc) on this admission and their complete blood count (cbc) was being monitored.There was an order for telemetry monitoring beginning (b)(6) at 08:50.The palliative team was consulted on this admission, and patient wished to remain full code.Philips received a complaint on the patient information center ix sn (b)(6) indicating that nursing leadership reviewed the telemetry monitor disclosure and it was identified that the patient went into asystole at 08:24 with pacer spikes firing.The customer alleged the central monitor did not alarm asystole until (b)(6) 2023 at 09:04.At 0905, a code blue was called after the patient was found unresponsive by the licensed practicing nurse (lpn).When reviewing the 20-minute telemetry disclosure, it was found that the patient had been asystole for 20 minutes prior to the code being called.The patient¿s pacemaker was spiking but there was no underlying rhythm.Advanced cardiac life support was initiated.Return of spontaneous circulation was unable to be achieved and the patient expired.The official cause of death was cardiac arrest.It was not unconfirmed whether the pacer detection was on or off at the time of the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT INFORMATION CENTER IX
Type of Device
PATIENT INFORMATION CENTER IX
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18400290
MDR Text Key331415205
Report Number1218950-2023-00967
Device Sequence Number1
Product Code MHX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K211900
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number866389
Device Catalogue Number866389
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 11/30/2023
Initial Date FDA Received12/27/2023
Supplement Dates Manufacturer Received01/04/2024
Supplement Dates FDA Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/28/2011
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Patient Outcome(s) Death;
Patient Age85 YR
Patient SexMale
-
-