C.R. BARD, INC. (BASD) -3006260740 POWER PORT ISP M.R.I., 8FR GROSHONG; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR Back to Search Results
Catalog Number 4808570J
Device Problems Restricted Flow rate (1248); Fracture (1260); Material Separation (1562)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the power port isp m.R.I., 8fr groshong that are cleared in the us.The pro code and 510 k number for the power port isp m.R.I., 8fr groshong are identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that sometime post a port placement, the catheter was allegedly ruptured.Reportedly, the port system was removed.There was no reported patient injury.
 
Manufacturer Narrative
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the power port isp m.R.I., 8fr groshong that are cleared in the us.The pro code and 510 k number for the power port isp m.R.I., 8fr groshong are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one powerport implantable port attached to a groshong catheter was returned for evaluation.Functional, gross visual, tactile and microscopic visual evaluations was performed.A complete circumferential break was noted on the distal and proximal end of the attached catheter that was elliptical in shape and got separated.The distal catheter segment was received and measured approximately 11.2cm in length.The edges of the both the distal and proximal break end were noted to be uneven and the surface was noted to be granular.Upon infusion thrombus, was noted exiting with water.Aspiration was attempted and was successful.Therefore, the investigation is confirmed for reported fracture and identified material separation issue.However, the investigation is inconclusive for the reported flow issue as no objective evidence to confirm this was provided for review.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H10: b5, g3, h6 (device).H11: h6 (method, result, conclusion).H11: section a through f ¿ the information provided by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that sometimes post port placement procedure, the catheter was allegedly ruptured.It was further reported that the flow of the chemical solution was slightly worse than before.Reportedly, the port system was removed.There was no reported patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POWER PORT ISP M.R.I., 8FR GROSHONG
Type of Device
PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
brett curtice
800 w. rio salado pkwy
tempe, AZ 85281
4803032689
MDR Report Key18410064
MDR Text Key331511853
Report Number3006260740-2023-05900
Device Sequence Number1
Product Code LJT
UDI-Device Identifier00801741175725
UDI-Public(01)00801741175725
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K063377
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/04/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4808570J
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received03/28/2024
Supplement Dates FDA Received04/06/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Age60 YR
Patient SexFemale
-
-