BIOSENSE WEBSTER INC QDOT-MICRO, BI-DIRECTIONAL, F-J CURVE, C3, SPLIT HANDLE; CATHETER, PERCUTANEOUS, CARDIAC ABLATION, FOR TREATMENT OF ATRIAL FIBRILLATION
|
Back to Search Results |
|
Catalog Number D139504 |
Device Problems
Material Puncture/Hole (1504); Failure to Sense (1559)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that a patient underwent an atrial fibrillation (afib) ablation procedure with a qdot-micro, bi-directional, f-j curve, c3, split handle and post procedure the bwi product analysis lab identified a hole in the pebax.During the procedure, the carto 3 system was displaying a force sensor error 106 when the catheter was plugged into the patient interface unit (piu) and inserted into the body.The cable was replaced without resolution.The catheter was replaced and the issue was resolved.The procedure was continued.No patient consequences were reported.
|
|
Manufacturer Narrative
|
The product investigation was completed.Device evaluation details: the qdot micro device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis of the returned sample revealed reddish material and a hole in the pebax.The magnetic and force features were tested and no errors were observed.The force values and the vector were observed within specifications.No force issues were observed.The blood found inside the pebax area may contribute to the force sensor error reported.The root cause of the pebax damaged could be related to the handling of the device during the procedure however this cannot be conclusively determined.A manufacturing record evaluation was performed for the finished device 31099008l number, and no internal actions was found during the review.The issue reported by the customer was confirmed.The instructions for use contain the following information that should be considered: to ensure accurate force readings, verify that the force reading is near zero when the catheter is not in contact with tissue.If the force reading is not near zero when the catheter is not in contact with tissue, perform zeroing.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
|
|
Search Alerts/Recalls
|
|
|