The reported complaint that the autopulse platform (sn (b)(6) consistently displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message was confirmed during functional testing and review of the archive data.The root cause of the (ua)07 was due to failed load cell #1.The cracked cover and load cell failure were likely attributed to mishandling such as a drop, or the age of the platform.The autopulse platform was manufactured in 2017 and is over 7 years old, past its expected service life of 5 years.During visual inspection, unrelated to the reported complaint, a damaged front enclosure was observed.The observed physical damage appeared to be the characteristics of mishandling.The front enclosure was replaced to address the issue.A review of the archive data showed (ua) 07 error messages; thus, confirming the reported complaint.The platform failed initial functional testing due to the (ua) 07 advisory message displayed upon powering up the platform; thus, confirming the reported complaint.The over-reporting single point load cell #1 was replaced to remedy the complaint.Following service, the platform passed the load cell characterization check.The autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint and there were no similar complaints reported for autopulse with serial number (b)(6).
|