Model Number MODEL 100
Device Problem Visual Prompts will not Clear (2281)
Patient Problem Insufficient Information (4580)
Event Date 12/05/2023
Event Type  malfunction  
Event Description
The customer reported the autopulse platform (sn 33887r) consistently displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message.It is unknown when the problem occurred.However, patient use information was requested but no additional information was provided.
 
Manufacturer Narrative
Zoll has not yet received the autopulse platform in the complaint for investigation.A supplemental report will be filed if and when the product is returned and investigation has been completed.
 
Manufacturer Narrative
The reported complaint that the autopulse platform (sn (b)(6) consistently displayed user advisory (ua) 07 (discrepancy between load 1 and load 2 too large) error message was confirmed during functional testing and review of the archive data.The root cause of the (ua)07 was due to failed load cell #1.The cracked cover and load cell failure were likely attributed to mishandling such as a drop, or the age of the platform.The autopulse platform was manufactured in 2017 and is over 7 years old, past its expected service life of 5 years.During visual inspection, unrelated to the reported complaint, a damaged front enclosure was observed.The observed physical damage appeared to be the characteristics of mishandling.The front enclosure was replaced to address the issue.A review of the archive data showed (ua) 07 error messages; thus, confirming the reported complaint.The platform failed initial functional testing due to the (ua) 07 advisory message displayed upon powering up the platform; thus, confirming the reported complaint.The over-reporting single point load cell #1 was replaced to remedy the complaint.Following service, the platform passed the load cell characterization check.The autopulse platform passed the run-in test without any fault or error.The autopulse platform passed the final testing without any fault or error.Historical complaints were reviewed for service information related to the reported complaint and there were no similar complaints reported for autopulse with serial number (b)(6).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AUTOPULSE® RESUSCITATION MODEL 100
Type of Device
CARDIAC CHEST COMPRESSOR
Manufacturer (Section D)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer (Section G)
ZOLL CIRCULATION INC
2000 ringwood ave
san jose CA 95131
Manufacturer Contact
kim thoa nguyen
2000 ringwood ave
san jose, CA 95131
4084192922
MDR Report Key18411564
MDR Text Key331623126
Report Number3010617000-2023-01082
Device Sequence Number1
Product Code DRM
UDI-Device Identifier00849111000772
UDI-Public00849111000772
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K112998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,User Facility,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 02/05/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberMODEL 100
Device Catalogue Number8700-0730-01-66
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/05/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received01/29/2024
Supplement Dates FDA Received02/05/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/19/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-