• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BIOSENSE WEBSTER INC THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER; CARDIAC ABLATION PERCUTANEOUS CATHETER Back to Search Results
Catalog Number D134701
Device Problems Material Puncture/Hole (1504); Failure to Sense (1559)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/05/2023
Event Type  malfunction  
Manufacturer Narrative
The product investigation was completed.Device evaluation details: the smart touch unidirectional sf device was returned to biosense webster (bwi) for evaluation.A visual inspection and screening test of the returned device were performed following bwi procedures.Visual analysis revealed that there was a reddish material inside the pebax and a rupture on the surface of the tip area.The magnetic and force feature were tested, and no errors were observed.The force values and the vector were observed within specifications.No force issues were observed.However, the rupture at the pebax could be related to the issue reported by de customer.The root cause of the rupture at the pebax cannot be determined; however, based on the information available, the condition observed most likely was originated in someplace external to the manufacturing environment.All units are inspected prior to leaving the facility as there are functional tests and inspections at control points based on the process flow diagram (pfd) per its part number to avoid this type of damage from leaving the facility.A manufacturing record evaluation was performed for the finished device 31072187l number, and no internal actions related to the reported complaint condition were identified.The issue reported by the customer was confirmed.It should be noted that product failure is multifactorial.The instructions for use (ifu) contain the following recommendations: to ensure accurate force readings, verify that the force reading is near zero when the catheter is not in contact with tissue.If the force reading is not near zero when the catheter is not in contact with tissue, perform zeroing.If the force problem persists, replace the catheter cable or the catheter.In order to prevent damage to the catheter tip, use the insertion tube supplied with the catheter to advance or retract the catheter through the hemostasis valve of the sheath.After insertion, slide the insertion tube back toward the handle.As part of biosense webster's quality process, all devices are manufactured, inspected, and released to approved specifications.If additional information is received regarding this event, a supplemental 3500a report will be submitted to the fda.Manufacturer's reference number: (b)(4).
 
Event Description
It was reported that a patient underwent an atrial fibrillation (afib) paroxysmal cardiac ablation procedure with a thermocool® smart touch® sf uni-directional navigation catheter and post procedure the bwi product analysis lab identified a rupture on the surface of the pebax tip.During the procedure, a force sensor error (code unknown) appeared on the carto 3 system while the stsf catheter was connected.The cable was reseated and exchanged without resolution.The stsf catheter was exchanged and the issues resolved.The case continued successfully.No patient consequences were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
THERMOCOOL® SMART TOUCH® SF UNI-DIRECTIONAL NAVIGATION CATHETER
Type of Device
CARDIAC ABLATION PERCUTANEOUS CATHETER
Manufacturer (Section D)
BIOSENSE WEBSTER INC
31 technology drive, suite 200
irvine CA 92618
Manufacturer (Section G)
BIOSENSE WEBSTER INC (JUAREZ)
circuito interior norte
1820parque industrial salvacar
juarez 32599
MX   32599
Manufacturer Contact
kate karberg
31 technology dr
irvine, CA 92618
3035526892
MDR Report Key18414863
MDR Text Key331716182
Report Number2029046-2023-03121
Device Sequence Number1
Product Code LPB
UDI-Device Identifier10846835009774
UDI-Public10846835009774
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other
Type of Report Initial
Report Date 12/28/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberD134701
Device Lot Number31072187L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2023
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/28/2023
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/21/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
CARTO 3 SYSTEM.
-
-