WRIGHT MEDICAL TECHNOLOGY INC INFINITY ADAPTIS TIB SZ1 INFINITY ADAPTIS; PROSTHESIS, ANKLE, SEMI-CONSTRAINED, CEMENTED, METAL/POLYMER
|
Back to Search Results |
|
Catalog Number 33680001 |
Device Problems
Loss of Osseointegration (2408); Osseointegration Problem (3003); Migration (4003)
|
Patient Problem
Inadequate Osseointegration (2646)
|
Event Date 11/14/2023 |
Event Type
Injury
|
Manufacturer Narrative
|
Based on the available information the device will not be returned therefore an evaluation of the device cannot be performed.Should additional information become available, it will be provided in a supplemental report.H3 other text : device disposition unknown.
|
|
Event Description
|
As reported: "subsidence of the tibial component seen on in office x-ray.Patient was sent for ct scan for further evaluation.Actions taken: oral antibiotics, double upright brace locked in neutral.".
|
|
Event Description
|
As reported: "subsidence of the tibial component seen on in office x-ray.Patient was sent for ct scan for further evaluation.Actions taken: oral antibiotics, double upright brace locked in neutral.".
|
|
Manufacturer Narrative
|
Please note correction to b1 and h6.The reported event could not be confirmed, since the device was not returned for evaluation and no other evidences were provided.The device inspection was not possible as the product was not returned for investigation.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.More detailed information about the complaint event as well as the affected device must be available in order to determine the root cause of the complaint event.If device is returned or any further information is provided, the investigation report will be reassessed.
|
|