BAXTER HEALTHCARE CORPORATION EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP; SET, I.V. FLUID TRANSFER Back to Search Results
Catalog Number H938737
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Event Description
It was reported that a 250ml eva (ethyl vinyl acetate) tpn (total parenteral nutrition) bag leaked from central port during filling.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
 
Manufacturer Narrative
H4: the lot was manufactured from october 24, 2022 to october 25, 2022.H10: the device was received for evaluation.A visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed using tap water to fill the container, and a leak was observed at the spike port bonding area of the sample.The reported condition was verified.The cause of the condition could not be determined; however, the cause was most likely due to an inadequate or a lack of cyclohexanone application to the spike port connector when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EXACTAMIX TUBING SETS FOR BAXA EXACTA-MED PHARMACY PUMP
Type of Device
SET, I.V. FLUID TRANSFER
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
AVAILMED
c. industrial lt 001 mz 105
no 20905 int a, col cd ind.
tijuana, baja california 22444
MX   22444
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18416349
MDR Text Key331581291
Report Number1416980-2023-06800
Device Sequence Number1
Product Code LHI
UDI-Device Identifier00085412477213
UDI-Public(01)00085412477213
Combination Product (y/n)N
Reporter Country CodeHK
PMA/PMN Number
K900585
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date 01/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberH938737
Device Lot Number60407093
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/06/2023
Initial Date FDA Received12/28/2023
Supplement Dates Manufacturer Received01/23/2024
Supplement Dates FDA Received01/25/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-