H4: the lot was manufactured from october 24, 2022 to october 25, 2022.H10: the device was received for evaluation.A visual inspection did not identify any abnormalities that could have contributed to the reported condition.Functional leak testing was performed using tap water to fill the container, and a leak was observed at the spike port bonding area of the sample.The reported condition was verified.The cause of the condition could not be determined; however, the cause was most likely due to an inadequate or a lack of cyclohexanone application to the spike port connector when it was inserted to the spike port during the manufacturing process.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|