Model Number TABLETOP-JAPAN
Device Problem Energy Output Problem (1431)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Event Description
A physician reported that during the cataract surgery an ophthalmic system encountered a problem with its laser functionality, rendering it unusable.To proceed with the surgery, they replaced the console with another one and patient harm was not reported.
 
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Manufacturer Narrative
Service history was reviewed for the system.No service record relevant to the complaint reported event was found.However, the system was last serviced prior to the reported event per service record (sr) opened.The system found to meet all cosmetic and performance standards.Based on the information obtained, the root cause of the reported event is inconclusive.A manufacturing device history record (dhr) review was performed prior to product release to ensure that the product was manufactured in compliance with the device master record.Based on the assessment, the product met release criteria.A non-conformance based review of the serial number was performed and a potential contributing factor to the reported complaint was identified.A review for complaints reported against this serial number was performed.No similar complaints were reported for the product serial under investigation.Service was not requested as the reported event was resolved by rebooting the system.Based on the information obtained, the root cause of the reported event is inconclusive.Manufacturer will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION VISION SYSTEM
Type of Device
UNIT, PHACOFRAGMENTATION
Manufacturer (Section D)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer (Section G)
ALCON RESEARCH, LLC - IRVINE TECHNOLOGY CENTER
15800 alton parkway
irvine CA 92618
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8007579780
MDR Report Key18417217
MDR Text Key331615041
Report Number2028159-2023-01732
Device Sequence Number1
Product Code HQC
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K101285
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 02/18/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberTABLETOP-JAPAN
Device Catalogue Number8065751817
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received01/31/2024
Supplement Dates FDA Received02/18/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured10/12/2015
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
CONSTELLATION SURGICAL PROCEDURE PACK.
-
-