• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/27/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device was found to be contaminated with mycobacterium chimaera during periodic water sampling tests.There is no report of patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in middlesbrough, united kingdom.Livanova initiated an investigation.This report was due on november 29, 2023; however, due to a network disruption at livanova, the ability to submit mdrs was lost on november 19, 2023, before this report was ready to submit.The report was prepared and submitted immediately following restoration of the livanova systems.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: through follow-up communication under previous complaint from the same customer, livanova learned that the device is cleaned regularly per instructions for use and it is placed both outside and inside the operating theater during use.When placed inside, the fan is directed opposite to the patient at an estimated distance of 2,5 meters from the surgery field.In addition, it was confirmed that hospital does not use patient re-usable blankets, the h2o2 level is checked daily as per device instructions for use and the device is stored full when unused.When stored full, h202 check is daily performed during working days and sporadically performed during week-end.Moreover, a disposable pall-aquasafe water filter with an 0.2 m membrane or equivalent performance filter is used for tap water.Prior to initial operation and prior to storing the heater-cooler, the surfaces and water circuits are disinfected as per device instruction for use.Finally, h2o2 is regularly added when water in the tanks is changed and 3t aerosol collection set is replaced twice weekly after the allowed use period.Based on the information collected, it cannot be excluded the above mentioned deviation from device instructions for use (missed h202 level check during weekend, when the device is stored full of water) may have led/contributed to the reported contamination.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich 80939
GM   80939
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key18417460
MDR Text Key331585850
Report Number9611109-2023-00667
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup
Report Date 03/25/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/30/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received02/27/2024
Supplement Dates FDA Received03/25/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured07/14/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-