Model Number 16-02-80
Device Problem Microbial Contamination of Device (2303)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Event Description
Livanova deutschland received a report that a heater-cooler system 3t device was found to be microbiologically contaminated during water sampling test.There is no report of patient injury.
 
Manufacturer Narrative
A.1.-a.5.There was no patient involvement.G.5.The heater-cooler 16-02-80 is not distributed in the usa and it is similar to heater-cooler 16-02-85, which is distributed in the usa (510(k) number: k191402).H.9.Livanova deutschland implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.H10: livanova deutschland manufactures the heater-cooler system 3t.The incident occurred in hasselt, belgium.Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Manufacturer Narrative
The device is not cleaned after every use and this is not in accordance with ifu's.Therefore, it cannot be excluded that highlighted deviation from ifu's may have contributed to reported contamination.Livanova has implemented a strategy to progressively decrease the probability of bacteria growth in the hc device by applying multiple measures implemented over the past few years through dedicated capa and field action.The risk is in the acceptable region.No specific action was currently deemed necessary.Livanova will keep monitoring the market.
 
Event Description
See initial report.
 
Manufacturer Narrative
Through follow up, livanova was informed the hospital does not use reusable blankets, the water quality monitoring is applied and the h2o2 checked every day as prescribed by the ifu, the device is used almost everyday except the weekend, device is stored full and the h2o2 is checked daily even during days of non-use (i.E.Week-end), h2o2 is added when the water in the tanks is changed (each 7 days), a disposable pall-aquasafe water filter with an 0.2 m membrane is used for tap water or equivalent performance filter, prior to initial operation and prior to storing the heater-cooler, the surfaces and water circuits are disinfected, device surfaces are not disinfected after every operation and are disinfected once a week (this is not in line with product ifu), the 3t aerosol collection set is replaced after the allowed use period, the device is placed outside of the operation theatre during use.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstrasse 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w 65th way
arvada, CO 80004
MDR Report Key18417464
MDR Text Key331689354
Report Number9611109-2023-00664
Device Sequence Number1
Product Code DWC
UDI-Device Identifier04033817901099
UDI-Public(01)04033817901099(11)170801
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 04/10/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-80
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received01/22/2024
03/16/2024
Supplement Dates FDA Received02/20/2024
04/11/2024
Was Device Evaluated by Manufacturer? No
Date Device Manufactured08/01/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
-
-