Model Number N/A |
Device Problem
Fracture (1260)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported that the tip of the guide rod broke while being tightened to the impactor.There was no known impact or consequences to the patient.It was reported that no further information is available.
|
|
Manufacturer Narrative
|
(b)(4).The customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
|
|
Manufacturer Narrative
|
(b)(4).This follow-up report is being submitted to relay additional information.Visual examination of the returned product identified the device has fractured at the tip.There are wear lines on the shaft and damage near under the fracture site.Complaint confirmed based on evaluation of the returned product.Review of the device history record(s) identified no deviations or anomalies during manufacturing.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further information is available at the time of this report.
|
|