It was reported to philips that the allura system did not start.The device was outside of clinical use at the time of the reported event, no harm was reported to philips.As per the field service engineer, the host pc was faulty.Philips has started an investigation of this complaint.
|
Philips has investigated this complaint.A philips field service engineer (fse) inspected the system on-site.The system log files were analyzed which did not indicate a malfunction, however as the host pc is responsible for maintaining the logging and it was alleged that the host pc would not boot, this is expected.As part of troubleshooting, all plugs on the host pc were checked and hard disks were reinserted.The fse successfully restarted the system without error.However, due to the intermittent start-up problem, the fse replaced the host pc and hard disk as a precautionary measure, reloaded the software and configured the new pc.The cause of the host pc failure could not be conclusively determined by the supplier¿s part analysis; however, the replacement of host pc by the fse has resolved the reported issue and restored the functionality of the system.After replacement of the host pc, the system was returned to use in good working order.
|