PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. ALLURA XPER FD; SYSTEM, X-RAY, ANGIOGRAPHIC Back to Search Results
Model Number ALLURA XPER FD20
Device Problem Activation Failure (3270)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/20/2023
Event Type  malfunction  
Event Description
It was reported to philips that the allura system did not start.The device was outside of clinical use at the time of the reported event, no harm was reported to philips.As per the field service engineer, the host pc was faulty.Philips has started an investigation of this complaint.
 
Manufacturer Narrative
Philips has investigated this complaint.A philips field service engineer (fse) inspected the system on-site.The system log files were analyzed which did not indicate a malfunction, however as the host pc is responsible for maintaining the logging and it was alleged that the host pc would not boot, this is expected.As part of troubleshooting, all plugs on the host pc were checked and hard disks were reinserted.The fse successfully restarted the system without error.However, due to the intermittent start-up problem, the fse replaced the host pc and hard disk as a precautionary measure, reloaded the software and configured the new pc.The cause of the host pc failure could not be conclusively determined by the supplier¿s part analysis; however, the replacement of host pc by the fse has resolved the reported issue and restored the functionality of the system.After replacement of the host pc, the system was returned to use in good working order.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ALLURA XPER FD
Type of Device
SYSTEM, X-RAY, ANGIOGRAPHIC
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL  5684 PC
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
veenpluis 6
best 5684 PC
NL   5684 PC
Manufacturer Contact
derrick massuri
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18418191
MDR Text Key331587951
Report Number3003768277-2023-07264
Device Sequence Number1
Product Code IZI
UDI-Device Identifier00884838059054
UDI-Public00884838059054
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K130842
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/15/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberALLURA XPER FD20
Device Catalogue Number722012
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/20/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received06/26/2024
Supplement Dates FDA Received07/15/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-