C.R. BARD, INC. (BASD) -3006260740 POWER PORT ISP M.R.I., 8FR GROSHONG; PORT & CATHETER, IMPLANTED, SUBCUTANEOUS, INTRAVASCULAR
|
Back to Search Results |
|
Catalog Number 4808570J |
Device Problems
Fracture (1260); Deformation Due to Compressive Stress (2889); Naturally Worn (2988)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/12/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the power port isp m.R.I., 8fr groshong that are cleared in the us.The pro code and 510 k number for the power port isp m.R.I., 8fr groshong are identified in d2 and g4.H10: as the lot number for the device was not provided, a review of the device history records could not be performed.The return of the sample is pending.The investigation of the reported event is currently underway.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that approximately two years post port placement procedure, the catheter allegedly had a break.Reportedly, the port was removed and replaced.There was no reported patient injury.
|
|
Manufacturer Narrative
|
H10: the catalog number identified in section d4 has not been cleared in the us but is similar to the power port isp m.R.I., 8fr groshong that are cleared in the us.The pro code and 510 k number for the power port isp m.R.I., 8fr groshong are identified in d2 and g4.H10: manufacturing review: a manufacturing review was not requested as the lot number reported is unknown.Investigation summary: one powerport implantable port attached to a groshong catheter was returned for evaluation.Visual, microscopic, tactile and functional evaluations were performed on the returned device.Two partial circumferential break was noted from the cath-lock and also a split was noted.The edges of the partial circumferential breaks on the attached catheter were noted to be uneven and the surfaces were noted to be round and smooth on both regions.Therefore the investigation is confirmed for the reported fracture and the identified naturally worn and deformation issues.The definitive root cause could not be determined based upon available information.Labeling review: as the reported event did not allege a labeling or use related issue, a labeling review is not required.H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|
|
Event Description
|
It was reported that approximately two years post port placement procedure, the catheter allegedly had a break.Reportedly, the port was removed and replaced.There was no reported patient injury.
|
|