Catalog Number 955606
Device Problem Inadequate Ultra Filtration (1656)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/04/2023
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported during treatment with an ak 98 machine, a high ultrafiltration (uf) occurred.The expected uf was 450ml; however, it was observed that the machine removed 525ml.There was no report of patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information added to h3, h6, and h10: h10: the device was evaluated on-site by a qualified technician.The qualified technician observed that the machine removed 525 ml instead of removing 450 ml.It was 25 ml more than the acceptable criteria of +-50ml; therefore, it was determined that the ultrafiltration (uf) measuring unit was faulty.To resolve the issue, the qualified technician replaced the uf measuring unit.A simulated treatment was performed after replacing the uf measuring unit, and it was noted that the uf was within acceptable range.The reported condition was verified.The cause of the reported condition could not be determined.A service history review revealed no indication that the parts replaced during servicing caused or contributed to the reported event.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AK 98 MACHINES
Type of Device
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MEDOLLA MODENA MONITORS
via modenese, 66
medolla modena 41036
IT   41036
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key18420839
MDR Text Key331634470
Report Number9616240-2023-00014
Device Sequence Number1
Product Code KDI
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 02/01/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number955606
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/04/2023
Initial Date FDA Received12/29/2023
Supplement Dates Manufacturer Received01/18/2024
Supplement Dates FDA Received02/01/2024
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-