The customer reported to olympus, the evis exera iii xenon light source would not read 180 and 190 scopes.The issue was found during preparation for use.The patient and equipment were moved to another room, which created a 10 minute delay.The diagnostic procedure was completed without issue.The device was returned for evaluation.During the device evaluation, the device was found to have no image.There were no reports of patient harm.This medical device report (mdr) is being submitted to capture the reportable malfunction found during evaluation.
|
The device was returned to olympus for evaluation and the customer's allegation was not confirmed.The device evaluation found the housing front panel mouth was cracked, the top cover was rusted, the front panel chassis was rusted, the bottom chassis was rusted, the lamp door was rusted, the connections socket slider switch was worn out causing intermittent use of the high intensity mode, corrosion was present inside the scope socket tubing, the non-olympus lamp life meter reading was under 50 hours, the light intensity measured out of range, the lamp was missing three washers, the non-olympus lamp upon inspection was observed to have damage thread-holes, and corrosion was present inside the air pump tubing.Based on the results of the investigation, it is likely that the following led to the malfunction: estimated cause of the 180 scope image failure was a failure of the cv-190 patient board.190 scope image failure was not reproduced during the inspection of any of the devices, so the cause could not be estimated.A definitive root cause cannot be identified.The non-olympus lamps were used for less than 50 hours and the light intensity was out of range.All three lamp washers were not attached and the screw holes in the lamps were broken during inspection.A device history review confirmed that the devices were shipped in compliance with the specifications.This issue is addressed in the instructions for use (ifu): the clv-190 instruction manual (figure no.Gt7243) contains the following description, which may prevent the indicated phenomenon.[6.1 replacement of the examination (xenon) lamp].Never install a lamp that has not been approved by olympus.The use of a nonapproved lamp can cause damage to the light source and ancillary equipment, malfunction or a fire.[6.2 removal of the lamp].Washers are attached to bolts.Retain all three of them carefully because they will be used again to clamp the new examination lamp.Olympus will continue to monitor the field performance of this device.
|