The customer reported to olympus, the cysto-nephro videoscope tested positive for an unexpected contamination.The issue was found during routine microbiological culture of the scope.All channels were sampled.The user did not report any contamination or any other serious deterioration in state of health of any person, to which the scope could have been a contributory cause.
|
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, a definitive root cause could not be determined.Growth of microorganisms were found through culture testing by the user after reprocessing, however, when olympus culture tested after reprocessing in accordance with the instructions for use (ifu) before repair, the results conformed to the regulation's recommendation.As a result of checking cds check list provided by the customer, it was confirmed that there was no obvious flaw in the reprocessing of the device; the device cleaning/reprocessing was found to be in accordance with the ifu, however, the evaluation noted that the inner wall of the forceps channel was scraped by a treatment instrument and there were shavings as a result of the scraping.It is possible that the observed shavings may have prevented thorough cleaning of the device and the reported microbial issue was the result of insufficient device cleaning/reprocessing.The issue may be prevented by following the instructions for use section below: cysto-nephro videoscope olympus cyf-vh olympus cyf-vhr reprocessing manual olympus will continue to monitor field performance for this device.
|