• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (BASD) -3006260740 NIAGARA SLIMCATH DIALYSIS CATHETER KIT 12F X 20CM STRAIGHT SHORT-TERM DUAL-LUMEN; CATHETER, HEMODIALYSIS, NON-IMPLANTED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (BASD) -3006260740 NIAGARA SLIMCATH DIALYSIS CATHETER KIT 12F X 20CM STRAIGHT SHORT-TERM DUAL-LUMEN; CATHETER, HEMODIALYSIS, NON-IMPLANTED Back to Search Results
Model Number N/A
Device Problem Crack (1135)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Event Description
It was reported the puncture needle is ruptured at the end of the needle, and the puncture and catheterization cannot be successfully completed.No other information was provided.
 
Manufacturer Narrative
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
 
Manufacturer Narrative
H11: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Based on a review of this information, the following was concluded: the complaint of a leaking/damaged needle hub is confirmed; however, the exact cause is unknown.Three photographs of an 18 g introducer needle were returned for evaluation.An initial visual observation of the photographs showed a relatively large crack in the hub of the introducer needle.Use residue was observed on the sample in the returned photographs.While a crack in the hub of the introducer needle can be seen, it was not possible to determine a root cause of this damage from the returned photographs.Therefore, this complaint is confirmed as ¿cause unknown¿ at this time.This complaint will be recorded for future trending and monitoring purposes.
 
Event Description
It was reported the puncture needle is ruptured at the end of the needle, and the puncture and catheterization cannot be successfully completed.No other information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NIAGARA SLIMCATH DIALYSIS CATHETER KIT 12F X 20CM STRAIGHT SHORT-TERM DUAL-LUMEN
Type of Device
CATHETER, HEMODIALYSIS, NON-IMPLANTED
Manufacturer (Section D)
C.R. BARD, INC. (BASD) -3006260740
605 north 5600 west
salt lake city 84116
Manufacturer (Section G)
BARD REYNOSA S.A. DE C.V. -9617592
blvd. montebello #1
parque industrial colonial
reynosa, tamaulipas 88780
MX   88780
Manufacturer Contact
maddy vincent
605 north 5600 west
salt lake city 84116
8015950700
MDR Report Key18430358
MDR Text Key332607994
Report Number3006260740-2023-05975
Device Sequence Number1
Product Code MPB
UDI-Device Identifier00801741045363
UDI-Public(01)00801741045363
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K010778
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number5553200
Device Lot NumberREHR2999
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/13/2023
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received01/02/2024
Supplement Dates Manufacturer Received01/26/2024
Supplement Dates FDA Received02/21/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/01/2023
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-