C.R. BARD, INC. (BASD) -3006260740 NIAGARA SLIMCATH DIALYSIS CATHETER KIT 12F X 20CM STRAIGHT SHORT-TERM DUAL-LUMEN; CATHETER, HEMODIALYSIS, NON-IMPLANTED
|
Back to Search Results |
|
Model Number N/A |
Device Problem
Crack (1135)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/06/2023 |
Event Type
malfunction
|
Event Description
|
It was reported the puncture needle is ruptured at the end of the needle, and the puncture and catheterization cannot be successfully completed.No other information was provided.
|
|
Manufacturer Narrative
|
H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The manufacturer has received the sample and will evaluate.Results are expected soon.
|
|
Manufacturer Narrative
|
H11: the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.Based on a review of this information, the following was concluded: the complaint of a leaking/damaged needle hub is confirmed; however, the exact cause is unknown.Three photographs of an 18 g introducer needle were returned for evaluation.An initial visual observation of the photographs showed a relatively large crack in the hub of the introducer needle.Use residue was observed on the sample in the returned photographs.While a crack in the hub of the introducer needle can be seen, it was not possible to determine a root cause of this damage from the returned photographs.Therefore, this complaint is confirmed as ¿cause unknown¿ at this time.This complaint will be recorded for future trending and monitoring purposes.
|
|
Event Description
|
It was reported the puncture needle is ruptured at the end of the needle, and the puncture and catheterization cannot be successfully completed.No other information was provided.
|
|