DEPUY INTERNATIONAL LTD - 8010379 ASR UNI FEMORAL IMPL SIZE 49; METAL FEMORAL HEAD Back to Search Results
Catalog Number 999890249
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Necrosis (1971); Pain (1994); Distress (2329); Discomfort (2330); Deformity/ Disfigurement (2360); Swelling/ Edema (4577)
Event Date 07/17/2023
Event Type  Injury  
Event Description
On (b)(6) 2023, plaintiff suffered the following personal and economic injuries as a result of the implantation with the asr hip implant: severe pain and discomfort, increased metal levels in blood including cobalt and chromium; permanent injuries, emotional distress, disability, disfigurement.Doi: (b)(6) 2007; dor: (b)(6) 2023; right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).D4: the device catalog number is unknown; therefore, udi is unavailable.Investigation summary: no device associated with this report was received for examination.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot: device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.E3 initial reporter occupation: lawyer.H6: component code: appropriate term/code not available (g07002) used to capture no findings available.This report is being submitted pursuant to the provisions of 21 cfr, part 803 (and/or part 4, as applicable).This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information that has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event Description
Medical records received.The patient had a right hip failed total hip replacement secondary to trunnionosis with pseudotumor.The patient developed pain in (b)(6) 2023.Esr was 30 and crp was <3.0 mg/l.Her metal ions were about 16 and 8 concerning for trunnionosis.Mars mri shows significant fluid collection.Findings were about 30ml of dark fluid was removed and black residue at the base of the trunnion.Necrotic tissue within the back of the femoral head and extending inferiorly.Debrided the necrotic tissue around the acetabulum.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The asr platform was voluntarily recalled from the market in august 2010, and the asr product codes are now considered inactive.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.Additional complaint information monitoring for potential safety signals will be conducted through complaint trending as part of the post-market surveillance.If additional information is made available, the investigation will be updated as applicable.Device history lot device history reviews for asr platform have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR UNI FEMORAL IMPL SIZE 49
Type of Device
METAL FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
Manufacturer (Section G)
LEEDS MFG & MATERIAL WAREHOUSE
st anthonys road
leeds
Manufacturer Contact
kate karberg
700 orthopaedic dr.
warsaw, IN 46581-0988
3035526892
MDR Report Key18430649
MDR Text Key331861801
Report Number1818910-2024-00141
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 01/02/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/12/2011
Device Catalogue Number999890249
Device Lot Number2250251
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/21/2023
Initial Date FDA Received01/02/2024
Supplement Dates Manufacturer Received01/10/2024
01/17/2024
Supplement Dates FDA Received01/16/2024
01/18/2024
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/12/2006
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR CUP ASR; UNK HIP FEMORAL SLEEVE ASR; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age78 YR
Patient SexFemale
-
-