1.Dhr/bhr review (lot#2110889): 1) this batch of products were assembled at intima ii auto line 2 in may 2022, and packaged at cfs package line in may 2022.Work order quantity was (b)(4) ea.2) review the in-process test reports and outgoing test reports, and all test results meet the product specifications.3) review the production records with no nonconformance, deviation or rework activities.2.No defective samples and photos have been received for the complaint.3.Check the unit packages of the retained samples of the complaint batch, no poor seal is found.Please see attachment for the inspection report.4.No similar complaints have been received from other hospitals regarding this batch of products.Conclusion(s): no abnormality is found on process and retained samples, and no similar complaints have been received from other hospitals about this batch of products.As the poor seal state of the unit package cannot be identified, the root cause of this defect cannot be determined.
|