U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC MINIMED UNKNOWN CARELINK; GLUCOSE OXIDASE, GLUCOSE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC MINIMED UNKNOWN CARELINK; GLUCOSE OXIDASE, GLUCOSE Back to Search Results
Model Number UNK_CARELINKSOFTWARE
Device Problem Computer Software Problem (1112)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/06/2023
Event Type  malfunction  
Manufacturer Narrative
Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.No conclusion can be drawn at this time.We therefore consider this report complete to the best of our knowledge.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Select patient information cannot be provided due to regional privacy regulations.
 
Event Description
Information received by medtronic indicated that the customer reported issue in uploading data through carelink.No harm requiring medical intervention was reported.Troubleshooting was performed.Product will not be returned for analysis.
 
Manufacturer Narrative
"complaint summary: user reported error message stating ""unable to upload data to server"".Investigation/testing summary: an attempt to reproduce the issue was not performed as it related to a hardware issue with user's pump.Carelink support team investigated carelink uploader database logs from time frame where user reported seeing the error message.Confirmed there were teneo errors present in related logs.Carelink support team confirmed with teneo team that pump's status in teneo was not fully configured, and that pump kept failing to complete registration steps.The software successfully adhered to the specified requirements and performed in accordance with the expectations specified in the (b)(4), version pch00113968.(most likely) root cause: pump unable to complete teneo registration process.Analysis summary: teneo team recommended replacement.Carelink support team advised helpline of recommendation.Helpline stated they would manage replacement of pump.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN CARELINK
Type of Device
GLUCOSE OXIDASE, GLUCOSE
Manufacturer (Section D)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325
Manufacturer (Section G)
MEDTRONIC MINIMED
18000 devonshire st.
northridge CA 91325
Manufacturer Contact
lusine boyadzhyan
18000 devonshire st.
northridge, CA 91325
MDR Report Key18434067
MDR Text Key331772900
Report Number2032227-2024-101123
Device Sequence Number1
Product Code CGA
Combination Product (y/n)N
Reporter Country CodeAS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Consumer
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date 02/26/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberUNK_CARELINKSOFTWARE
Device Catalogue NumberUNK_CARELINKSOFTWARE
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 12/06/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received12/18/2023
Supplement Dates FDA Received02/26/2024
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage A
Patient Sequence Number1
Patient Age49 YR
-
-