Model Number 799-AEB1-00-01 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Event Description
|
Complainant alleged that during a routine shift check by a clinician, the device displayed an "internal comm failure -1472" error message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
|
|
Manufacturer Narrative
|
The customer's report was observed during review of the device logs.However, the device was put through extensive testing including bench handling, power cycling, and functional stress testing without duplicating the report.An internal inspection resulted in no findings.The 13 pin central processing unit (cpu) to power interface module (pim) ribbon cable was replaced as precaution.The device was recertified and returned to the customer.Analysis of reports of this type has not identified an increase in trend.
|
|