Model Number 101967-710
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Myalgia (2238); Discomfort (2330); Swelling/ Edema (4577)
Event Date 02/22/2021
Event Type  Injury  
Manufacturer Narrative
This report is being submitted retrospectively as part of an internal review.A review of the device's manufacturing records revealed that the sensor lot met all requirements including sterilization requirements for release.The sensor is inserted by making a small incision and placing it under skin and potential for developing skin irritation/pain/ inflammation/infection at the insertion site is a known anticipated adverse event.It was last reported that the sensor will be removed on (b)(6).But no further information was received despite making several attempts.
 
Event Description
On (b)(6) 2021, senseonics was made aware of an incident where patient experienced pain at the insertion site after using the sensor for almost 6 months.Removal date was scheduled in the beginning of march.The patient felt pain for the last three days when insertion site is touched or when moving arm.It was a bit swollen but no redness.Patient said she will go see a doctor and if necessary removal date will be moved forward.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVERSENSE SENSOR
Type of Device
EVERSENSE CONTINUOUS GLUCOSE MONITOR SYSTEM
Manufacturer (Section D)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer (Section G)
SENSEONICS INC.
20451 seneca meadows parkway
germantown MD 20876 7005
Manufacturer Contact
vallikannu somasundaram
20451 seneca meadows parkway
germantown, MD 20876-7005
MDR Report Key18435819
MDR Text Key331859344
Report Number3009862700-2023-00564
Device Sequence Number1
Product Code QCD
Combination Product (y/n)Y
Reporter Country CodeAU
PMA/PMN Number
P160048
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial
Report Date 01/03/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Expiration Date10/25/2020
Device Model Number101967-710
Device Catalogue NumberFG-4400-10-302
Device Lot NumberWP07653
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/09/2021
Initial Date FDA Received01/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-