Catalog Number MCS20
Device Problems Failure to Form Staple (2579); Failure to Fire (2610); Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/01/2023
Event Type  malfunction  
Event Description
It was reported that during an unknown procedure, clips not working while clipping.No patient consequences.
 
Manufacturer Narrative
(b)(4); date sent: 1/3/2024.B3: unknown, assumed first day of month that complaint was reported.D4: batch # unk.Additional information was requested and the following was obtained: ""please clarify how ¿clips not working while clipping¿ did device not feed clips? - yea.Did device feed clips sideways? - not properly.Did device not fire clips (jammed)? - yes.Did device fire malformed clips?- yes.Did device fire scissored clips? - not properly.Did device drop or eject clips? - clips jumps out.If other, please specify." an analysis of the product could not be performed since a physical sample was not received for evaluation.However, if the product is received at a later date, the investigation will be updated as applicable.A manufacturing record evaluation was performed for the finished device lot number, and no non-conformance were identified.As part of our company quality system process, all devices are manufactured, inspected, and distributed to approved specifications.This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by ethicon, or its employees that the report constitutes an admission that the product, ethicon, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
(b)(4).Date sent: 2/13/2024.D4: batch # a9d67d.Investigation summar: the product was returned to ethicon for evaluation.Visual inspection and functional testing were conducted on the returned device.Visual analysis of the returned sample revealed that the mcs20 device was returned with no apparent damage.In an attempt to replicate the reported incident, the instrument was cycled and in the next actuations, no clips were fed into the jaws even though the device was being actuated in several occasions.The device was disassembled in order to evaluate the condition of the internal components and the feeder shoe was noted be damaged.Seven(7) clips were found inside clip track.No conclusion could be reached as to what may have caused the reported incident.The reported complaint was confirmed.As part of ethicon¿s quality process, all devices are manufactured, inspected, and released to approved specifications.A manufacturing record evaluation was performed for the finished device batch and lot number, and no non-conformances were identified.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer (Section G)
ETHICON ENDO-SURGERY, LLC.
475 calle c
guaynabo 00969
Manufacturer Contact
kate karberg
475 calle c
guaynabo 
3035526892
MDR Report Key18435966
MDR Text Key331901765
Report Number3005075853-2024-00036
Device Sequence Number1
Product Code GDO
UDI-Device Identifier10705036002499
UDI-Public10705036002499
Combination Product (y/n)N
Reporter Country CodeIN
PMA/PMN Number
K820837
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other
Type of Report Initial
Report Date 02/13/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberMCS20
Device Lot Number491C27
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 12/06/2023
Initial Date FDA Received01/03/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/07/2023
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
-
-