• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PHILIPS MEDICAL SYSTEMS SURESIGNS VS4 NBP, SPO2 Back to Search Results
Model Number 863283
Device Problems Device Alarm System (1012); No Audible Alarm (1019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/07/2023
Event Type  malfunction  
Manufacturer Narrative
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Reporter phone not provided.
 
Event Description
The customer reported audio failure with the system.It is unknown if the device was in use at time of event, and there was no adverse event reported.
 
Manufacturer Narrative
A philips field service engineer (fse) went onsite to evaluate the device in question.The fse indicated during an evaluation of the system that the system speaker and mainboard failed.The philips field service engineer (fse) replaced the system speaker and mainboard.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.
 
Event Description
The customer reported audio failure with the system.The device was in use on a patient at the time of event, there was no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SURESIGNS VS4 NBP, SPO2
Type of Device
SURESIGNS VS4 NBP, SPO2
Manufacturer (Section D)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer (Section G)
PHILIPS MEDICAL SYSTEMS
3000 minuteman rd
andover MA 01810
Manufacturer Contact
hisham alzayat
3000 minuteman rd
andover, MA 01810
6172455900
MDR Report Key18436758
MDR Text Key331809641
Report Number1218950-2024-00004
Device Sequence Number1
Product Code DSJ
UDI-Device Identifier00884838087095
UDI-Public00884838087095
Combination Product (y/n)N
Reporter Country CodeLU
PMA/PMN Number
K163649
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,User Facility
Reporter Occupation Biomedical Engineer
Type of Report Initial,Followup
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number863283
Device Catalogue Number863283
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 12/07/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received01/02/2024
Supplement Dates FDA Received01/23/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/16/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-