Model Number 863283 |
Device Problems
Device Alarm System (1012); No Audible Alarm (1019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 12/07/2023 |
Event Type
malfunction
|
Manufacturer Narrative
|
Philips is in the process of obtaining additional information concerning this event and the complaint is still under investigation.A final report will be submitted once the investigation is complete.Reporter phone not provided.
|
|
Event Description
|
The customer reported audio failure with the system.It is unknown if the device was in use at time of event, and there was no adverse event reported.
|
|
Manufacturer Narrative
|
A philips field service engineer (fse) went onsite to evaluate the device in question.The fse indicated during an evaluation of the system that the system speaker and mainboard failed.The philips field service engineer (fse) replaced the system speaker and mainboard.The device was operational after repairs were completed and the device was returned to the customer.The investigation concludes that no further action is required at this time.
|
|
Event Description
|
The customer reported audio failure with the system.The device was in use on a patient at the time of event, there was no adverse event reported.
|
|