ZIMMER BIOMET, INC. SCREWDRIVER HEX BALL END 3.5MM; INSTRUMENT, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 12/13/2023
Event Type  malfunction  
Event Description
It was reported that the screwdriver head broke off in the femoral stem.There was no consequences or impact to the patient.
 
Manufacturer Narrative
(b)(4).The product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product found that the found it to exhibit signs of repeated use.The hex feature of component has fractured off in the post screw component of the stem.The device history record was reviewed and no discrepancies relevant to the reported event were found.Medical records were not provided.The root cause of the reported issue is attributed to the over 11 years of use in the field.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SCREWDRIVER HEX BALL END 3.5MM
Type of Device
INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer (Section G)
ZIMMER BIOMET, INC.
56 e. bell drive
p.o. box 587
warsaw IN 46581
Manufacturer Contact
jennifer rapsavage
56 e. bell dr.
warsaw, IN 46582
5745260384
MDR Report Key18439179
MDR Text Key331886492
Report Number0001825034-2024-00021
Device Sequence Number1
Product Code HXX
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Health Professional,Distributor
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/16/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number32-467201
Device Lot Number350270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/22/2023
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 12/13/2023
Initial Date FDA Received01/03/2024
Supplement Dates Manufacturer Received05/15/2024
Supplement Dates FDA Received05/17/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2012
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Age70 YR
Patient SexMale
-
-